Comparison BIPM.RI(I)-K8 of high dose-rate Ir-192 brachytherapy standards for reference air kerma rate of the VSL and the BIPM

J.T. Alvarez, J.A. de Pooter, C. Andersen, A.H.L. Aalbers, P.J. Allisy-Roberts, C Kessler, “Comparison BIPM.RI(I)-K8 of high dose-rate Ir-192 brachytherapy standards for reference air kerma rate of the VSL and the BIPM”, Metrologia 51, Tech. Suppl. 06022 (2014).

Abstract:
An indirect comparison of the standards for reference air kerma rate for 192Ir high dose rate brachytherapy sources of the Dutch Metrology Institute (VSL), The Netherlands, and of the Bureau International des Poids et Mesures (BIPM) was carried out at the VSL in November 2009. The comparison result, based on the calibration coefficients for a transfer standard and expressed as a ratio of the VSL and the BIPM standards for reference air kerma rate, is 0.9873 with a combined standard uncertainty of 0.0061.Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).